Usage tips:卡式盒的使用技巧
1.Always note the position of the cassette when inserting it into unit-if the cassette is installed upsidedown,the thermocouple will bend and the sterilizer will become inoperative.
2.Lubricate the exposed area of the cassette seal with RPI Liquid Soap (RPI Part # RRS287) on a weekly basis
3.Alway keep the inside of the casstte clean.Remove residue and/or debris from the metal edge of the lower half of the cassette on a regular basis.
4.Watch for steam leaks from the cassette as they can cause damage to the microswith ,thus rendering the sterilizer inoprative.
5.Refill the reservoir with distilled water only.At the same time that the reservior is refilled.the condenser waste bottle should be emptied and refilled.Always refill the condenser waste bottle with tap water to the Minimum Fill Level as shown on the bottle(Note:Never operate sterilizer when water level is above the Maximum Fill Line or below the Minnum Fill Line)
2000 Replacement Parts Industries ,Inc,SCS001 INS REV B (08/04)